xHobo®
Multi-Channel Re-usable Data Loggers
|
|
|
|
|
|
|
|
Record temperature, humidity,
light intensity, and external inputs for under $100!
|
|
|
xTiny
xenough
xto fit xanywhere!
|
|
HO8-004-02
|
HO8-007-02
|
HO8-003-02
|
HO8-002-02
|
HO8-001-02
|
|
|
xx4
Channels:
|
xx4
Channels:
|
xx2
Channels:
|
xx2
Channels:
|
xx1
Channel:
|
|
|
xx-Temp
xx-Humidity
xx-Light
intensity xx-External
input
|
xx-Temp
xx-Humidity
xx-2
External xxxinputs
|
xx-Temp
xx-Humidity
|
xx-Temp
xx-External
input
|
xx-Temp
|
|
Use your PCx
to launch loggers and download data
|
|
xx1-5
@
$109 xx6+
@ $105
|
xx1-5
@
$109 xx6+
@
$105
|
xx1-5
@
$99 xx6+
@ $95
|
xx1-5
@
$79 xx6+
@
$75
|
xx1-5
@
$73 xx6+
@
$69
|
|
|
|
|
|
|
|
|
|
Software
kit required
|
|
Optional external
input sensors sold separately
|
|
|
|
|
|
|
Use
these match-book sized data loggers to collect and store temperature and
other readings around the clock! Launch the loggers from your
computer, place them in remote locations for desired length of time,
then download time/date-stamped data to your computer for graphing and
analysis. Hundreds of uses: hospitals, greenhouses,
laboratories, livestock facilities, warehouses, factories, schools,
museums, and more. Need more info on data-logging? Click
Here
Common
Features
- Capacity:
7943 measurements total
- User-selectable
sampling interval: 0.5 seconds to 9 hours, recording times up to 1
year
- Readout
and re-launch with optional HOBO
Shuttle
- Internal
temperature sensor on 4" wire can extend from case
- Models
with external input accept external sensors for temperature, AC
current, 4-20 mA and 0-2.5 Volts DC
- Precision
components eliminate the need for user calibration
- Drop-proof
to 5 feet
- Mounting
kit included (hook/loop, magnet, and tape)
- Programmable
start time/date
- Memory
modes: stop when full or wrap-around when full
- Nonvolatile
EEPROM memory retains data even if battery fails
- Blinking
LED light confirms operation
- 3-Volt
Lithium battery, included. User-replaceable.
- Battery
level indication at launch
- Operating
range: -4°F to +158°F
(-20°C to +70°C), 0 to 95% relative humidity, non-condensing,
non-fogging
- Time
accuracy: ±1 minute per week at +68°F (+20°C)
- Size/Weight:
2.4 x 1.9 x 0.8" (68 x 48 x 19 mm)/approx. 1 oz.(29 gms)
- A software starter kit is
required for operation
BoxCar Pro 4.0+
starter kit includes software, computer interface cable and software
manual.
|
|
Measurement
specifications
Temperature (internal sensor)
- Range: -4°F to +158°F (-20°C
to +70°C)
- Range for internal sensor when
used outside of case:
-40°F to +248°F
(-40°C to +120°C)
- Accuracy: ±1.27°F (±0.7°C)
at +70°F, see plot below
- Resolution: 0.7°F (0.4°C) at
+70°F
- Response time still in air: 15
min. typical with sensor inside case; 1 min. typical with sensor
outside case
Relative humidity
(user-replaceable RH sensor)
- Range:
25% to 95% RH at +80°F for intervals of > 10 seconds,
non-condensing and non-fogging, see plot below
- Accuracy:
±5%
- Response
time 10 min. typical in air
- Sensor
operating environment: +41°F to +122°F (+5°C to +50°C)
non-condensing and non-fogging
Light intensity
- Range:
2 to 600 footcandles (lumens/ft2)
- Accuracy:
±2 footcandles, ±20% of reading
- Angular
response: roughly cosine of the angle from directly above sensor
- Light
wavelength response, see plot below
External input
Loggers having external input capability accept
the following sensors:
- Wide
range of temperature sensors (TMCx-Hx series)
- Split-core
CTs for AC current (5 Models)
- 4-20
mA input cable (CABLE-4-20mA)
- 0-2.5
V DC input cable (CABLE-2.5-STEREO)
Optional external
input sensors sold separately
External 2.5 volt input specifications
- 2.5
mm jack: external input ground, input, switched 2.5 V output; external
input ground connection is not the same as PC interface connection
ground and should not be connected to any external ground
- Input
range: 0 to +2.5 Volts DC
- Accuracy:
±10 mV ±1% of reading
- Resolution:
10 mV (8-bit)
- Out
power: +2.5 Volts DC at 2 mA, active only during measurements
|
|